Swipe the Slider Bar to view all Six Menus
IS Adjustment Menu
Use this recalibration menu if the IS Unit or Main Board have been replaced, or if the IS Unit was removed during service.
- IS Shift is used to: Enter the shaft data (normally) provided with the new IS Unit.
- IS Gyro Rack is used to: Enter the gyro data (normally) supplied with the IS Unit or Gyro Sensor.
- IS Central Position is used to: Adjust to prevent the viewfinder from jumping when IS is turned On or Off.
- IS Vibration Adjust is used to: Fine-tune the IS Gain for optimal image stabilization.
Notes
- The IS Unit has several sub-assemblies, each requiring individual calibration through the IS Adjustment menus.
- In most cases, the unit comes factory-calibrated, and you simply enter the values provided with the new IS Unit.
- If calibration values are not supplied, the SPT software also allows manual adjustment.
IS Shift
Install your IS Unit… easily and quickly!
IS Unit Replacement Process:
Enter the Pitch and Yaw values from the IS Unit label or packaging,. then press ‘Write’. The software may ask you to remove and refit the lens to apply the new settings.
You should enter new IS Shift values when:
- Replacing or reinstalling the IS Unit
- Replacing the Lens Main Board
Accurate IS Shift data is essential for proper image stabilization.
⚠️ If you replace the Main Board without backing up the original data and don’t have the IS Shift values, you’ll need to replace the IS Unit.
The IS Shift data on the Main Board must match the values provided with the IS Unit.
IS Gyro Rack
Install your Gyro Rack… easily and quickly!
IS Gyro Rack Replacement Process:
Enter the Pitch and Yaw values from the Gyro Rack Unit label or packaging, then press ‘Write’.
IS Central Position
If the viewfinder image jumps when Image Stabilization is turned On or Off, use the IS Central Position menu to fix it.
Testing the IS Central Position:
- Press the shutter halfway or use the ‘Confirm Rest Position’ window in the software.
- If the image stays stable, no changes are needed.
- If it jumps or shifts, adjusting the IS Central Position can fix the issue.
Manual IS Center Adjustment:
Enter Pitch and Yaw values from the IS Unit label or packaging, then press ‘Write’.
Auto IS Center Adjustment:
With the camera on a tripod, click ‘Auto Adjust’. Use the resulting values in the Manual Adjustment window and press ‘Write’.
Confirm Rest Position:
Use the ‘Test’ button to ensure the image settles in the same spot each time. Repeat as needed.
IS Vibration Adjust
If the lens does not respond to camera shake or overcorrects, use the IS Vibration Adjust menu. Typically needed after replacing the IS Unit or Main Board.
Menu Sequence:
- IS Shift
- IS Gyro Calibration
- Test / Adjust IS Vibration
⚠️ The factory uses a calibrated vibration table, but most techs just rock the camera by hand. See Help files for more.
Testing by Hand
With IS On, point the camera at a small, high-contrast target. Move gently and observe the IS response.
Pitch Test:
Rock forward/back slowly, increasing to 3–4 times per second. Image should stay steady vertically.
Yaw Test:
Rock side-to-side, 3–6 times per second. Image should stay steady horizontally.
Adjusting Pitch and Yaw:
- Pitch: Adjust slider in steps of 50.
- Yaw: Rotate camera vertically, adjust slider, test in both directions.
- Once stable, press ‘Write’ to save.
⬅️ Done? Use your browser’s back arrow to return.











